background image

 

            ZEBV-5084A

 

2.3 Ambient requirements

 

 
This  instrument  is  intended  for  precision  measurement.

 

Sufficient  consideration  must  be  given  to 

preparing  its  installation  environment  as  a  given  level  of  performance  can  not  be  guaranteed  if 
installation conditions are not met. 
 
Temperature 

Constant temperature of 18 to 28°C (at all times, both day and night) 

 

Please  ensure  temperature  does  not  fluctuate  during  the  day  or  night.  This  will 
ensure maintenance of unit precision. Place unit away from direct sunlight and in a 
location where it will not come into direct contact with airflows from air conditioners. 

 

 

 

 

   Temperature stability: 

 

  + - 1.5° C    (Term - less than 1° C / 1 Hour) 

 

  

  + - 0.5° C    (Term - more than 1° C / 1 Hour) 

 

  

  + - 1.5° C    (Term – Rapid change within10minutes cycle ) 

 

  

 

 

   (Reference) Heat discharge of this instrument: around 2,000 kcal/h 

 
Humidity 

40 to 70% (with no condensation) 

 
Install location 

This instrument weighs 900 kg. Install it on a flat floor that can withstand a weight 
of 900 kg or more.  Ensure  that  there  is  around  2.7  m  ×  1.6  m  of  space  for 
installation  and  that  the  room  can  withstand  the  weight  of  the  instrument.  The 
position for installation of this instrument must be at least 2.3 m high. Including 
space for working, provide a W4 m 

 D3 m 

 H2.3 m site. 

 
Ventilation 

This instrument uses flammable solvents for auto tuning, and uses large amounts of

 

nitrogen gas. Therefore, using the instrument in an insufficiently ventilated room 
may cause oxygen deficiency. Prepare equipment that provides sufficient ventilation. 

 

 

Exhaust 

Be sure to install duct equipment such as a draft chamber for the exhaust gas from 
the rotary pump. 

 
Other 

• Install in a room with minimal dust. This instrument uses a high voltage, and 
dust may cause electrical discharge, damage or malfunction. A buildup of dust in the 
fan also may cause overheating. 

 

•  To  maintain  high  performance,  install  in  a  room  with  minimal  disturbing 
elements such as vibration, electromagnetic wave noise and corrosive gases. 

 

• Be sure to read “Safety Instructions” of the instruction manual before using this 
instrument. 

 
 

2.4 Required power supply 

 

 

For safety reasons use a power supply equipped with a ground fault interrupter.   

 

The iMScope TRIO unit uses AC230V, 20A, 50/60Hz (single phase) and  AC100 V, 15 A, 50/60 Hz 
(single phase). Avoid sharing power source with other equipment. 

 

A cable terminal supplying AC230V to the iMScope TRIO must be connected to a circuit breaker or to 
a power outlet using 3-point power plug. (Refer Fig3.) 

Содержание iMScope TRIO

Страница 1: ...ZEBV 5084A Imaging Mass Microscope iMScope TRIO Pre Installation Requirements Installation Preparations Confirmation...

Страница 2: ...on Copyright 2014 Shimadzu Corporation 2 Information in this manual is subject to change without notice and does not represent a commitment on the part of the vendor 3 Any errors or omissions which ma...

Страница 3: ...t requirements 6 2 4 Required power supply 6 2 5 Gas 7 2 5 1 Nitrogen gas 7 2 5 2 Argon gas 9 2 5 3 General precautions 9 2 5 4 Precautions when using high pressure gas cylinder 10 2 6 Exhaust 10 2 7...

Страница 4: ...reparation required at the customer s site in order to facilitate installation of the iMScope TRIO Please make sure that the content of this document is carefully observed to ensure that the instrumen...

Страница 5: ...crane prepare an approach route and the stopping place of the crane and the carry in entrance to the building Passage and the corners in the building should have the sufficient width and height to mov...

Страница 6: ...space for maintenance Place equipment that can be readily moved such as the PC system given in the example to the left of the iMScope TRIO unit and ensure a space of 60cm or more between the iMScope...

Страница 7: ...on of this instrument must be at least 2 3 m high Including space for working provide a W4 m D3 m H2 3 m site Ventilation This instrument uses flammable solvents for auto tuning and uses large amounts...

Страница 8: ...refers to the Shimadzu Corporation part number If power supplies of the above specifications are not available alterations to room power supply capacity will be necessary Required power supplies for e...

Страница 9: ...es and pressures Gas Pressure supplied to iMScope TRIO Type Purity Nitrogen gas N2 99 or greater In case of nitrogen gas generator 97 0 or greater 690 800kPa Argon gas A 99 99 or greater 200 500kPa 2...

Страница 10: ...Prepare the following pressure regulator and pipes given below or equivalent products if using a nitrogen gas cylinder Pressure regulator Part number Description Requirement Remarks 221 35999 01 Press...

Страница 11: ...en supplying gas by centered gas supply as other instruments being used may be affected by the amount of gas supplied If using a cylinder please use 5m or shorter pipes in general carrier gas piping I...

Страница 12: ...e Draft chamber 20m3 min Be sure to install the oil return kit to suppress the decrease in oil in the rotary pump and clean the exhaust gas from a rotary pump The parts which are require for the exhau...

Страница 13: ...a draft chamber during the process of spray coating of the matrix to prevent dispersion of the spray solution of the matrix Dustproof Gasproof mask We recommend you use a Dustproof Gasproof mask duri...

Страница 14: ...ce For measuring the matrix Pipette and pipette tips 100 1000uL Matrix solvent HPLC grade methanol This is also used for cleaning airbrush Milli Q ultrapure water TFA Vortex mixer For agitating the ma...

Страница 15: ...the following pages are a checklist for each of the installation conditions that have been described in this document Please fill out the checklist yourself and pass it on to the appropriate personnel...

Страница 16: ...Screen to avoid direct contact with airflows from air conditioners Humidity 40 to 70 with no condensation Installation space Space for iMScope TRIO Strength Portable table for PC and monitor Stand fo...

Страница 17: ...e situation 3 point power plug for 230V Power strip for PC Monitor 4 Gas N2 nitrogen gas Purity Supply pressure to iMScope TRIO Distance from gas supply source 99 or greater In case of nitrogen gas ge...

Страница 18: ...us Adjustment acceptance Standard samples for instrument adjustment and sensitivity check are required CD R Blank for installation data record is required TFANa 200mL P N 225 06613 08 Standard sample...

Страница 19: ...cable Strait 2pc Others Consumable parts Multimeter is required to distinguish which is the right side of the ITO coated glass slide ITO coated microscope slides and air duster are consumed regularly...

Страница 20: ...hen affixing the sample Because of the instrument composition we can not analyze that part c Section Thickness To avoid the section damage and the instrument trouble we ask you to slice the tissue wit...

Страница 21: ...al Cutting Temperature compound causes sensitivity decline because the polymer adherents to the sections If you cannot avoid using OCT we ask you to make sections outside the OCT as well If you wish t...

Отзывы: