9 |
P a g e
INTRODUCTION
Note:
A CO
2
gas regulator must be purchased separately from the incubator.
CO
2
G
AS
S
UPPLY
Supply Required
The incubator must be connected to a carbon dioxide gas supply system in order to
establish and maintain a CO
2
-enriched incubation chamber atmosphere. The supply
can be a building CO
2
gas system or one or two supply cylinders (tanks).
Supply Quality
Use medical or food grade CO
2
. Use of industrial CO
2
risks introducing contaminants into the chamber,
may damage the incubator, and will void the manufacturing defect warranty.
Supply Source and Pressure
The incubator requires 15 - 20 psi of CO
2
gas pressure at the incubator intake port (labeled CO
2
to
Chamber).
Two Stage Regulator
If connecting to a supply cylinder,
always use a two-stage CO
2
pressure regulator
. During normal
operations, the incubator uses small quantities of CO
2
to maintain the chamber gas concentration.
Under these conditions, precise regulation of the gas input flow is vital for the incubator
performance. Some single-stage regulators have two gauges. Make certain your regulator is a two-
stage regulator.
Regulator Not Included
See page 60 for ordering information.
Dual Cylinder Connection.
The incubator can be connected simultaneously to two CO
2
supply cylinders. The unit
automatically switches drawing from one cylinder to the other when the first cylinder
runs out of CO
2
. Two gas regulators are required for this configuration.
Содержание SCO5W
Страница 1: ...WATER JACKETED CO2 INCUBATOR 110 120 Volts Installation Operation Manual SCO5W ...
Страница 2: ...2 P a g e ...
Страница 6: ...6 P a g e TABLE OF CONTENTS ...
Страница 14: ...14 P a g e RECEIVING YOUR UNIT ...
Страница 24: ...24 P a g e INSTALLATION ...
Страница 56: ...56 P a g e USER MAINTENANCE ...
Страница 61: ...61 P a g e PARTS LIST ...