53 |
P a g e
USER MAINTENANCE
Continued next page
CO
2
Calibration Continued
3
Place the incubator CO
2
display in its calibration mode.
a.
Press and hold both the
UP
and
DOWN
temperature arrow buttons simultaneously for
approximately 5 seconds.
b.
Release the buttons when the temperature display
shows the letters “C O”. The display will then
begin flashing the current CO
2
display value.
Note:
If an arrow key is not pressed for five seconds, the
display will cease flashing, and store the last displayed
number as the CO
2
display value.
4
Use the
Up
or
Down
arrows to adjust the current display
CO
2
% value until it matches the reference device
concentration reading.
Reference Device
5
After matching the display to the reference device, wait 5
seconds.
•
The gas display will cease flashing and store
the corrected chamber display value.
•
The incubator will now begin injecting CO
2
or
allowing the current concentration to decay in
order to achieve the set point with the
corrected display value.
Decaying to Set Point
6
Allow the incubator sit for at last one 1 hour undisturbed to
stabilize
after the incubator has achieved the corrected
gas set point
.
•
Failure to wait until the incubator is fully
stabilized will result in an inaccurate reading.
Set Point Achieved
Содержание SCO5W
Страница 1: ...WATER JACKETED CO2 INCUBATOR 110 120 Volts Installation Operation Manual SCO5W ...
Страница 2: ...2 P a g e ...
Страница 6: ...6 P a g e TABLE OF CONTENTS ...
Страница 14: ...14 P a g e RECEIVING YOUR UNIT ...
Страница 24: ...24 P a g e INSTALLATION ...
Страница 56: ...56 P a g e USER MAINTENANCE ...
Страница 61: ...61 P a g e PARTS LIST ...