45 |
P a g e
USER MAINTENANCE
G
AS
L
INES AND
HEPA
F
ILTERS
For the Optional Filtered Tubing Kit
The manufacturer recommends replacing in-line gas HEPA filters once per year or
when a filter is noticeably discolored.
•
HEPA filters are directional and must be installed facing in the correct direction.
•
The word “IN” is stamped on the rim of the filter assembly on the side that faces
toward the gas supply. See page 22.
Gas lines should be replaced when cracking, brittleness, permanent kinking, or other
signs of damage are present.
S
TORAGE OF THE
I
NCUBATOR
Perform the following steps if the incubator will be out of use for more than 24 hours to prevent
microbiological contamination such as fungus or mold.
Depower the incubator.
Disinfect and clean if required by your laboratory protocol, or if the chamber has been
exposed to pathogenic microorganisms.
Use a soft cloth to dry the chamber surfaces.
•
Do not place the incubator into storage while the chamber surfaces are damp.
M
AINTAINING
A
TMOSPHERIC
I
NTEGRITY
Periodically, inspect the door latch, trim, catch, and gaskets for signs of deterioration. Failure to
maintain the integrity of the door system shortens the life span of the incubator.
E
LECTRICAL
C
OMPONENTS
Electrical components do not require maintenance. If the incubator fails to operate as specified,
please contact your distributor or
for assistance (please see page 8).
“IN”
In-Line Gas Filter
Содержание SCO5W
Страница 1: ...WATER JACKETED CO2 INCUBATOR 110 120 Volts Installation Operation Manual SCO5W ...
Страница 2: ...2 P a g e ...
Страница 6: ...6 P a g e TABLE OF CONTENTS ...
Страница 14: ...14 P a g e RECEIVING YOUR UNIT ...
Страница 24: ...24 P a g e INSTALLATION ...
Страница 56: ...56 P a g e USER MAINTENANCE ...
Страница 61: ...61 P a g e PARTS LIST ...