Audible and visual deviation alarms activate if the incubation chamber
temperature deviates by 1
°
C above or below the temperature set point.
The low deviation audible alarm has a delay of 15 minutes to prevent the
low audible from sounding each time the doors are opened, causing a
short-lived drop in temperature.
To mute an audible high or low deviation alarm:
Press and hold either the
Up
or
Down
arrow on the Temperature
Control until the amber Mute LED illuminates and the audible alarm
shuts off.
•
The audible alarm will remain muted for the duration of the
current temperature deviation.
•
The visual alarm indicator will remain illuminated.
•
Any
new
deviation of ±1°C or greater will reactivate the audible
alarm.
A
UTOMATIC
D
OOR
C
UTOFF
Gas, heating, and air circulation pause.
Whenever the incubation chamber door is opened, the incubator automatically stops the flow of CO
2
into the chamber, depowers the heater element, and ceases operation of the internal blower fan.
This limits the amount of CO
2
released into the workspace around the incubator. It also prevents the
heater from attempting to counteract the continual inflow of cooler air, which would cause a
significant heat spike once the door is closed. Normal CO
2
injections, heating, and fan operation all
resume automatically when the outer door is closed.
Figure 2: High
Alarm Muted
Содержание SCO5W
Страница 1: ...WATER JACKETED CO2 INCUBATOR 110 120 Volts Installation Operation Manual SCO5W ...
Страница 2: ...2 P a g e ...
Страница 6: ...6 P a g e TABLE OF CONTENTS ...
Страница 14: ...14 P a g e RECEIVING YOUR UNIT ...
Страница 24: ...24 P a g e INSTALLATION ...
Страница 56: ...56 P a g e USER MAINTENANCE ...
Страница 61: ...61 P a g e PARTS LIST ...