27 |
P a g e
CONTROL PANEL OVERVIEW
Control Panel
Power Switch
Power is supplied when the switch is in the ( I ) on position.
Set Over Temperature
This graduated dial sets the Over Temperature Limit system heating cut off point. The OTL system
helps prevents unchecked heating of the incubation chamber in the event the main temperature
control system fails. For more details, please see the
page 30.
The red Over Temp Activated light illuminates when the Over Temperature Limit system cuts off
heating to the incubation chamber by rerouting power away from the heating elements.
Temperature Control and Display
During normal operations, the Set Temperature display shows the current incubation chamber air
temperature, accurate to 0.1°C. The Up and Down buttons are used to change display modes and
then input either a new temperature set point or a calibration adjustment. The display blinks
continually while in its set point or calibration adjustment modes, preceded by an “SP” for Set Point
or “C O” for calibration offset.
Red LED alarm indicators marked High and Low illuminate whenever the temperature deviates by
±1°C or greater from the current set point. The yellow LED marked Mute illuminates whenever an
audible deviation alarm is being muted. See the
Muting the Audible Temperature Alarm
page 35 for more information.
The green indicator labeled Heating Activated illuminates whenever the temperature control system
is heating the incubation chamber.
Содержание SCO5W
Страница 1: ...WATER JACKETED CO2 INCUBATOR 110 120 Volts Installation Operation Manual SCO5W ...
Страница 2: ...2 P a g e ...
Страница 6: ...6 P a g e TABLE OF CONTENTS ...
Страница 14: ...14 P a g e RECEIVING YOUR UNIT ...
Страница 24: ...24 P a g e INSTALLATION ...
Страница 56: ...56 P a g e USER MAINTENANCE ...
Страница 61: ...61 P a g e PARTS LIST ...