17 |
P a g e
INSTALLATION
P
OWER
S
OURCE
R
EQUIREMENTS
When selecting a location for the unit, verify each of the following requirements is satisfied.
Power Source
The power source for the unit must match the voltage and match or exceed the amperage
requirements listed on the unit data plate. These units are intended for
110 - 120V 50/60 Hz
applications at
5 Amps
•
Supplied voltage must not vary more than 10% from the data plate rating
. Damage to the
unit may result if the supplied voltage varies more than 10%.
•
The wall power source must be protective earth grounded.
•
Use a separate circuit to prevent loss of the unit due to overloading or circuit failure.
•
The recommended wall circuit breakers for these units are 15 amps.
•
The wall power source must conform to all national and local electrical codes.
Power Cord
The unit must be positioned so that all users can quickly unplug the cord in the event of an
emergency.
•
Each unit comes provided with a
115 volt 15 Amp, 9.2ft (2.86m) NEMA 5-15P
power cord.
•
Always use this cord or an identical replacement.
Fuses
These units ship with a fuse installed in a fuse holder next to the power cord inlet.
•
The fuse must be installed and intact for the unit to operate.
•
Always find and fix the cause of a blown fuse prior to putting the unit back into operation.
•
Fuse type
o
250V, T20, 5X20mm
NEMA 5-15R
wall socket
Содержание SCO5W
Страница 1: ...WATER JACKETED CO2 INCUBATOR 110 120 Volts Installation Operation Manual SCO5W ...
Страница 2: ...2 P a g e ...
Страница 6: ...6 P a g e TABLE OF CONTENTS ...
Страница 14: ...14 P a g e RECEIVING YOUR UNIT ...
Страница 24: ...24 P a g e INSTALLATION ...
Страница 56: ...56 P a g e USER MAINTENANCE ...
Страница 61: ...61 P a g e PARTS LIST ...