OPERATION
Note:
Test the OTL system at least once per year for functionality.
S
ET THE
O
VER
T
EMPERATURE
L
IMIT
This procedure sets the OTL heating cutoff point to approximately 1
°
C above the current incubation
chamber temperature.
The unit must be operating stabilized at your application temperature for at
least 1 hour prior to setting the OTL cutoff system.
If the OTL is sporadically activating, you may turn the dial very slightly to the right (clockwise).
If the OTL continues activating, check for ambient sources of heat or cold that may be adversely
impacting the unit temperature stability. Check if any powered accessories in the chamber are
generating heat. If you find no sources of external or internal temperature fluctuations, contact Tech
Support or your distributor for assistance.
End of Procedure
1. Set OTL control to its maximum setting,
if not already set to max.
2.
Turn the dial counterclockwise until the red Over Temperature Limit Light illuminates.
3.
Slowly turn the dial clockwise until the OTL Activated light turns off.
•
The Over Temperature Limit is now set approximately 1˚C
above the current incubator air temperature.
4.
Leave the OTL dial set just above the activation point
.
Optional: Turn the dial slightly to the left.
•
This sets the OTL cutoff threshold nearer to the current
incubator air temperature.
Содержание SCO5W
Страница 1: ...WATER JACKETED CO2 INCUBATOR 110 120 Volts Installation Operation Manual SCO5W ...
Страница 2: ...2 P a g e ...
Страница 6: ...6 P a g e TABLE OF CONTENTS ...
Страница 14: ...14 P a g e RECEIVING YOUR UNIT ...
Страница 24: ...24 P a g e INSTALLATION ...
Страница 56: ...56 P a g e USER MAINTENANCE ...
Страница 61: ...61 P a g e PARTS LIST ...