USER MAINTENANCE
R
EPLACE THE
C
HAMBER
HEPA
F
ILTER
Always turn off and unplug the incubator before carrying out this procedure.
Replace the filter at least once per year, or when the filter is noticeably discolored.
Exercise caution
when removing the ceiling air duct. A plastic blower fan and the fragile head of the
temperature and CO
2
sensors are located just above the duct and just to the right of the duct.
Dismount the ceiling air duct, using firm but careful pressure to remove the duct feet from the
tops of the shelf standards.
Remove the duct from the chamber.
Remove the black plastic cap from the HEPA filter by pulling down on it.
Remove the old HEPA filter by pulling down. It will snap out without difficulty.
Snap the new HEPA filter into position on the duct. It may be necessary to tilt the filter slightly
to one side.
filter on page 20 for detailed instructions on
reinstalling the filter and duct.
Step 1
Step 2
Step 3
Step 4
Back
Front
Ceiling Duct
HEPA Filter with