USER MAINTENANCE
End of procedure
CO
2
Calibration Continued
7
Compare the reference device reading with the incubator CO
2
display again.
•
If the reference device and the incubator display readings
are the same or the difference falls within the range of your
protocol,
the incubator is now calibrated for CO
2
.
-OR-
•
See the next step if the readings fail to match or fall outside
of your protocol range.
Reference Device
8
If the two readings are not the same, and the difference still falls
outside the acceptable range of your protocol, repeat steps 3 – 7
up to two (2) more times.
•
Three (3) calibration attempts may be required to
successfully calibrate units that are more than ±2% out
of calibration.
Reference Device
9
Contact your distributor or technical support for assistance if the gas concentration readings
of the incubator CO
2
display and the reference device still fall outside your protocol after three
calibration attempts.