Separation of the rubber from the pins.
Swelled or bulging rubber pins.
6.
Apply a small quantity of 3M
™
Spray Silicone Lubricant (3M ID# 62-4678-4930-3) to
the pin end of the connector. Make sure to let it dry.
7.
Connect the connectors.
8.
Use a lint-free wipe to clean any unwanted lubricant from the sides of the connectors.
4.7 Long term storage
The user needs to do several steps to store the sensor for several weeks to several
months.
Note: Keep the sensor at or below 25 °C when possible. Temperatures above 40 °C cause an
increase in the sensitivity of the sensor. It takes several days for the sensor to return to normal
operation.
C A U T I O N
Bleach and Triton
®
X-100 are corrosive. Do not store the sensor with either solution still in the
sensor.
1.
Clean the sensor in fresh water. Refer to
CT-DO maintenance
on page 21 for details.
2.
Clean the BLIS and make sure there is no bleach in the sensor. Refer to
Examine
and maintain BLIS
on page 28 for details.
3.
Shake any excess water out of the sensor. If there is a possibility that the sensor
could freeze, store it where temperatures are above freezing and keep the sensor
warm before deployment.
4.
Cover the intake and the exhaust ports on the CT-DO guard with tape.
The calibration drift is reduced if air does not enter the sensor.
Sensor maintenance
32