Adjust the BLIS for the type of environment it will be deployed in.
•
BLIS Operation
—Set to "On." When the conductivity value shows that the sensor is
in water, the BLIS sends bleach to the pump manifold after the first sample of every
hour.
•
Purge when in situ
—Set to "Yes." When the sensor is in situ, the BLIS pump
operates 20 times to remove air or water from the BLIS tubing. This makes sure a
sufficient quantity of bleach is available to protect the DO sensor.
•
BLIS Interval
—The BLIS sends the selected quantity of bleach after the first sample
of every hour.
•
BLIS Injection Volume
—Select the quantity of bleach to be used in 9 µL increments.
The manufacturer recommends a minimum of 36 µL.
•
BLIS Volume Used
—The host software makes a continuous estimate of the quantity
of bleach used in relation to the quantity selected by the user.
•
Estimated BLIS Remaining
—The host software makes a continuous estimate of the
quantity of bleach in the reservoir. The estimate is in relation to the
Sample Interval
,
the
Injection Volume
, and the
BLIS Volume Used
.
5.1.5 Power use estimates
The software calculates the power required by the sensor to complete a deployment in
the 'New Setup' Power Consumption area.
Note: The user should add 10–25 % to the estimates of the software for factors such as battery
chemistry self-discharge.
The "Estimated Power Consumption @ 12 V" uses the values from the "Sample For:," the
"Interval Time:," and the "DO Stabilization:" time in relation to the user-selected expected
water temperature to estimate how much power the sensor will use during 24 hours of
data collection.
The total power use for a deployment is the daily power use multiplied by the number of
days the user selects in the "Deployment Days" slider.
5.1.6 Sensor internal disk storage
The Internal Disk Storage shows the condition of the memory in the sensor.
•
File #—Shows the current file number and the size of the file.
•
Free—Shows the available memory in the sensor.
•
Used—Shows the total size of all of the files stored in the sensor.
•
Total—Shows the total memory of the sensor.
The Free Disk at the bottom of the Setup Parameters column shows an estimate of the
number of days the sensor can store data before the memory is full.
Reference