4.2.1 Replace CT-DO intake anti-fouling cartridge
1.
Use a #2 Phillips screwdriver to remove the two screws on opposite sides of the
copper sleeve. Do not remove the three screws on the top.
Figure 3 CT-DO end cap removal
2.
Pull the end cap up and off.
Do not twist
.
3.
Use a #0 Phillips screwdriver to loosen the four screws that attach the duct top to the
base.
Figure 4 Anti-fouling duct screws
Note that the screws have size 003 O-rings under their heads.
4.
Lift the duct top straight up and off.
5.
View the anti-fouling cartridge in the duct base.
Figure 5 Anti-fouling cartridge
If necessary, push the size 021 O-ring back in its groove.
6.
Use a cotton swab to remove the anti-fouling cartridge from the duct base. To prevent
damage to the duct base, do not push the swab all the way through the cartridge.
Sensor maintenance
23