Figure 13 Wiper attached with set screw
7.
Remove the wiper height gauge.
4.4 Examine pressure sensor
C A U T I O N
Do not use a brush or any object in the pressure sensor. It is easy to break.
Examine the pressure sensor at regular intervals and remove any particles or debris.
Figure 14 Pressure sensor
4.5 Examine and maintain BLIS
C A U T I O N
Sediment and other solids will cause the BLIS intake area to become clogged and it will not
function correctly.
C A U T I O N
Do not use any abrasive the inside of the BLIS intake. Scratches on this surface will prevent the
nipple of the bleach reservoir from a proper seal.
The Bleach Injection System (BLIS) uses a Playtex Drop-Ins
®
liner as a bleach reservoir.
This product is made of a recyclable (check local with the local waste carrier) HDPE
plastic that can withstand long-term exposure to bleach. DO NOT substitute with another
liner.
Make sure that the BLIS operates correctly. Operate the system with water as a
operational check. Only fill the reservoir with bleach immediately before deployment.
1.
Follow the steps in
Set up BLIS for deployment
on page 18 to remove the bleach
reservoir.
2.
Use a 5/32" hex key to loosen the clamp that holds the CT sensors and BLIS.
Sensor maintenance
28