Section 2
Functional check
W A R N I N G
FDOM sensors use an ultraviolet LED light source. Do not look directly at a UV LED when it is on.
It can damage the eyes. Keep products that have UV LEDs away from children, pets, and other
living organisms. Wear polycarbonate UV-resistant safety glasses to protect the eyes when a UV
LED is on.
The WQM has configurations of sensors that use the abbreviations listed below.
•
WQM—Water Quality Monitor
•
BLIS—BLeach Injection System
•
CTD—Conductivity, Temperature, Depth
•
DO—Dissolved Oxygen
•
FDOM—Fluorescent Dissolved Organic Matter
Model
Description
WQM-a
Available analog output. BLIS anti-fouling. Chlorophyll fluorometer and turbidity sensor. CTD and DO sensors.
WQM-d
Digital output. BLIS anti-fouling. Chlorophyll fluorometer and turbidity sensor. CTD and DO sensors.
WQMx-a
Available analog output. BLIS anti-fouling. Chlorophyll fluorometer, turbidity, and CDOM sensors. CTD and DO
sensors.
WQMx-d
Digital output. BLIS anti-fouling. Chlorophyll fluorometer, turbidity, and CDOM sensors. CTD and DO sensors.
2.1 Pre-deployment check
The user needs the items listed below to make sure that the sensor operates correctly
before deployment.
•
a clean work bench
•
a PC
•
a regulated power supply
Manufacturer-supplied:
•
test cable
•
sensor
•
CD with software
•
5/32" hex key
•
orange fluorescent stick
•
blue fluorescent stick
1.
Lubricate each of the connectors with pure silicone spray. Let the spray dry. Refer to
the section on
"bulkhead connector maintenance"
for details about how to care for
bulkhead connectors.
2.
Select the software, the file on the CD that ends with ".exe" and move it to the
desktop of the PC.
The software is now available on the PC.
3.
Connect the test cable to the PC, the sensor, and the power supply.
4.
Start the software on the PC.
5.
Supply power to the sensor.
The software tries to identify the COM port and the baud rate.
6.
If necessary, select the COM port and the baud rate (19200) from the drop-down
menu in the COM Port Setup area of the software.