Section 5
Reference
5.1 Setup selections
The user can view or adjust the sensor configurations below in the
Setup
tab.
•
the internal clock
•
the sample parameters
•
the BLIS parameters
•
the condition of the BLIS
•
the quantity of data storage used
•
the condition of the Uninterruptible Power Supply (UPS)
5.1.1 Date and time
The software shows the date and the time for both the WQM and the PC. This information
has a red background if the clocks are different by more than 5 seconds.
Push
Set Date and Time
to update the date and time in the sensor to the PC.
5.1.2 Get and send setup information
The values that the user can change are listed under the Setup Parameters heading.
Push
Get Current Setup
to get the setup information that is stored in the sensor. This
information will show under the Current Setup heading.
Any setup values that the user changes will show under the New Setup heading and have
a yellow background. Push
Send Current Setup
to save new values to the sensor. The
yellow background goes out of view.
Note: Push Get Current Setup to make sure that the values are stored in the sensor.
5.1.3 Dissolved Oxygen (DO) stabilization time
Push
Show DO Stable vs. Temp. Plot
to see a graph of the relationship between the
temperature of the water and the time required to get a stable DO value.
5.1.4 Bleach Injection System (BLIS) operation
The BLIS sends bleach into the pump manifold just below the DO sensor. The bleach
prevents biological fouling on the DO sensor membrane.