![SCIEX 3200 QTRAP Скачать руководство пользователя страница 98](http://html1.mh-extra.com/html/sciex/3200-qtrap/3200-qtrap_system-users-manual_1217527098.webp)
Source Parameters and Voltages
D
TurboIonSpray Probe Parameters
The following table shows the recommended operating conditions for the TurboIonSpray probe
at three different flow rates. For each flow rate, the flow rate of the gas for the Curtain Gas
interface should be as high as possible. The solvent composition used for optimization was 1:1
water:acetonitrile. These conditions represent a starting point from which to optimize the probe.
Using an iterative process, optimize the parameters using flow injection analysis to achieve the
best signal or signal-to-noise for the compound of interest.
Table D-1 Parameter Optimization for the TurboIonSpray Probe
Parameters
Typical Values
Operational
Range
Low
Medium
High
LC flow rate
5 µL/min
to 50
µL/min
200 µL/min
1,000 µL/min
5 µL/min to
3,000 µL/min
Ion source gas 1
(nebulizer gas)
20 psi to
40 psi
40 psi to 60 psi
40 psi to 60 psi
0 psi to 90 psi
Ion source gas 2
(heater gas)
0 psi
50 psi
50 psi
0 psi to 90 psi
IonSpray Voltage
5500 V
5500 V
5500 V
5500 V
Gas for the Curtain Gas
interface
20 psi
20 psi
20 psi
20 psi to 50 psi
Ion source
temperature
Ambient to
200 ºC
200 ºC to 650 ºC 400 ºC to 750 ºC Up to 750 ºC
Declustering potential
(DP)
Positive:
70 V
Negative:
–70 V
Positive: 70 V
Negative: –70 V
Positive: 100 V
Negative: –100 V
Positive: 0 V to
400 V
Negative: –400 V
to 0 V
1
Optimum temperature values depend on the compound and mobile phase composition. Higher aqueous
content requires a higher temperature. Zero (0) means no temperature is applied.
2
DP values depend on the compound.
Source Parameters and Voltages
SCIEX 3200 Systems
98/114
System User Guide
RUO-IDV-05-6475-C