Calibration Ions and Solutions
A
CAUTION: Potential Wrong Result. Do not use expired solutions or solutions that have not been stored at the indicated
storage temperature.
Note:
Immediately after use, put a cap on the bottle and then store it at 2 °C to 8 °C. Refer to the information on the label.
Table A-1 Tuning Frequency
Calibration
Resolution Optimization
Scan Type
Frequency
Frequency
Q1 and Q3
3 months to 6 months
3 months to 6 months
LIT
3 months to 6 months
3 months to 6 months
Table A-2 Suggested Tuning Solutions for the 3200 Series of Instruments
System
Q1 and Q3
LIT
Positive
Negative
Positive and Negative
API 3200 LC-MS/MS system
POS PPG, 1e–5 M
NEG PPG, 3e–4 M
N/A
3200 QTRAP LC-MS/MS
system
POS PPG, 1e–5 M
NEG PPG, 3e–4 M
NEG PPG, 3e–4 M
Calibration Ions and Solutions
SCIEX 3200 Systems
82/114
System User Guide
RUO-IDV-05-6475-C