1. Wipe the external surfaces with a soft cloth dampened with warm, soapy water.
2. Wipe the external surfaces with a soft cloth dampened with water to remove any soap
residue.
Clean the Front-End
The following warning applies to all of the procedures in this section:
WARNING! Hot Surface Hazard. Let the Turbo V ion source cool for at least
30 minutes before starting any maintenance procedures. Some surfaces of
the ion source and vacuum interface become hot during operation.
Clean the mass spectrometer front-end using the routine cleaning method, to:
• Minimize unscheduled system downtime.
• Maintain optimum sensitivity.
• Avoid more extensive cleaning that requires a service visit.
When contamination occurs, perform an initial routine cleaning. Clean up to and including the
front of the orifice plate. If routine cleaning does not resolve issues with sensitivity, then a full
cleaning might be necessary.
This section provides instructions for performing routine cleaning without breaking vacuum.
Note:
Follow all of the applicable local regulations. For health and safety guidelines, refer to the
.
Symptoms of Contamination
The system might be contaminated if any one of the following is observed:
• Significant loss in sensitivity
• Increased background noise
• Additional peaks that are not part of the sample in full scan or survey scan methods
If any of these issues are observed, then clean the mass spectrometer front-end.