2. Wait five minutes, and then turn on the mass spectrometer convenience switch.
3. Turn on the computer.
4. Open the control software.
Syringe Pump
Adjust the Integrated Syringe Pump Position
WARNING! Puncture Hazard. Take care when handling the syringe. The tip of
the syringe is extremely sharp.
WARNING! Puncture Hazard. Make sure that the syringe is seated properly
in the syringe pump and that the automatic syringe pump stop is adjusted
properly to avoid damaging or breaking the glass syringe. If the syringe
breaks, follow established safety procedures for sharps disposal.
For the location of the syringe pump on the mass spectrometer, refer to the figure:
.
1. Press the Release button on the right side of the syringe pump to lower the base and then
insert the syringe.
2. Make sure that the end of the syringe is flush with the base and that the shaft of the syringe
rests in the cutout.
3. Adjust the post so that it triggers the automatic syringe stop before the syringe plunger hits
the bottom of the glass syringe.
4. Make sure that the mass spectrometer and integrated syringe pump are activated in the
software.
5. In the Analyst software, on the Navigation bar, double-click
Manual Tuning
.
6. Click
Start Syringe
.
7. To stop the syringe pump, click
Stop Syringe
.
Diverter Valve
To configure the valve, access the
Configuration
tab and then make sure that the
Use
integrated injector/diverter valve
check box is selected.