Replace the Sample Tubing
WARNING! Electrical Shock Hazard. Remove the ion source from the mass
spectrometer before starting this procedure. Follow all electrical safe work
practices.
Prerequisite Procedures
• Stop the sample flow and make sure that any remaining gas has been removed through the
source exhaust system.
• Remove the ion source. Refer to the section:
Use the following procedure to replace the sample tubing if it has a blockage.
1. Disconnect the sample tubing from the probe and the grounding union.
2. Replace the sample tubing with an appropriate length of tubing, cut with a proper tubing
cutter. Refer to the section:
3. Install the ion source. Refer to the section:
.
4. Start the sample flow.
Storage and Handling
WARNING! Environmental Hazard. Do not dispose of system components in
municipal waste. Follow local regulations when disposing of components.
The environmental requirements for the storage and transport of the ion source:
• Ambient temperature between –30 °C and +60 °C (–22 °F and 140 °F)
• Atmospheric pressure between 75 kPa to 101 kPa
• Relative humidity not exceeding 99%, non-condensing
Service and Maintenance — Ion Source
SCIEX 3200 Systems
76/114
System User Guide
RUO-IDV-05-6475-C