Performance Criteria A (Criteria A):
Equipment shall operate as intended with no degradation
of performance and no loss of function during or after test.
Performance Criteria B (Criteria B):
Equipment may experience loss of function (one or more)
during test but shall operate as intended with some degradation of performance and functions
self-recoverable after test.
Performance Criteria C (Criteria C):
Equipment may experience loss of function (one or more)
during test but shall operate as intended with some degradation of performance and functions
recoverable by operator after test.
The equipment is intended for use in a basic electromagnetic environment.
The expected performance loss under the electromagnetic immunity conditions is less than 20%
change in total ion count (TIC).
Make sure that a compatible electromagnetic environment for the equipment can be maintained
so that the device will operate as intended. If the power supply line is subject to high electrical
noise, then install a surge protector.
Electromagnetic Interference
Group 1 Equipment:
This equipment is classified as industrial, scientific, and medical (ISM)
equipment that might use RF energy for internal operation.
Class A Equipment:
Equipment which is suitable for use in all establishments other than
domestic and those directly connected to a low voltage power supply network which supplies
buildings used for domestic purposes. [Derived from CISPR 11:2009, 5.3] Class A equipment
shall meet Class A limits.
CAUTION: Potential Radio Interference. This equipment is not intended for use in
residential environments and may not provide adequate protection to radio reception in
such environments.
This equipment has been tested and found to comply with the limits for a Class A digital device,
pursuant to Part 15 of the FCC (Federal Communications Commission) Compliance Rules.
These limits are designed to provide reasonable protection against harmful interference when
the equipment is operated in a commercial environment. This equipment generates, uses,
and can radiate radio frequency energy and, if not installed and used in accordance with the
operator's manual, can cause harmful interference to radio communications.
Operation of this equipment in a residential area is likely to cause harmful interference in
which case you will be required to correct the interference, at your own expense. Changes or
modifications not expressly approved by the manufacturer could void your authority to operate
the equipment.