• Overall system run time
For information about tuning frequency, refer to the section:
Calibration Ions and Solutions
To order consumable parts and for basic service and maintenance requirements, contact a QMP
or refer to the document:
Parts and Equipment Guide
. Contact a SCIEX FSE for all other service
and maintenance requirements.
Table 7-1 Mass Spectrometer Maintenance Tasks
Component
Frequency
Task
For More Information
System
Daily
Inspect for leaks Refer to the section:
.
Curtain plate
Daily
Clean
.
Roughing pump oil
Weekly
Inspect the level Refer to the section:
Contact the local QMP or FSE to
add oil, if required.
Card cage air filter
Every 6
months
Replace
Contact the local QMP or FSE.
Roughing pump oil
Every 6
months, or as
needed
Replace
Contact the local QMP or FSE.
Roughing pump oil
As needed
Refill
Contact the local QMP or FSE.
Orifice plate (front)
As needed
Clean
Orifice plate (front and
back)
As needed
Clean
Contact the local QMP or FSE.
Skimmer
As needed
Clean
Contact the local QMP or FSE.
Instrument surfaces
As needed
Clean
.
Source exhaust drain
bottle
As needed
Empty
.
Interface heater
As needed
Replace
Contact the local QMP or FSE.
Service and Maintenance Information — Mass Spectrometer
SCIEX 3200 Systems
56/114
System User Guide
RUO-IDV-05-6475-C