Laboratory Conditions
Safe Environmental Conditions
The system is designed to operate safely under these conditions:
• Indoors
• Altitude: Up to 2,000 m (6,560 ft) above sea level
• Ambient temperature: 5 °C (41 °F) to 40 °C (104 °F)
• Relative humidity: 20% to 80%, non-condensing
• Mains supply voltage fluctuations: ± 10% of the nominal voltage
• Transient overvoltages: Up to the levels of Overvoltage Category II
• Temporary overvoltages on the mains supply
• Pollution Degree 2
Performance Specifications
The system is designed to meet specifications under these conditions:
• An ambient temperature of 15 °C to 30 °C (59 °F to 86 °F)
Over time, the temperature must remain within a range of 4 °C (7.2 °F), with the rate
of the change in temperature not exceeding 2 °C (3.6 °F) per hour. Ambient temperature
fluctuations exceeding the limits might result in mass shifts in spectra.
• Relative humidity from 20% to 80%, non-condensing
Equipment Use and Modification
WARNING! Personal Injury Hazard. Contact the SCIEX representative if product
installation, adjustment, or relocation is required.
WARNING! Electrical Shock Hazard. Do not remove the covers. Removing
the covers might cause injury or malfunctioning of the system. The covers
need not be removed for routine maintenance, inspection, or adjustment.
Contact a SCIEX Field Service Employee (FSE) for repairs that require the
covers to be removed.
Operational Precautions and Limitations
System User Guide
RUO-IDV-05-6475-C
SCIEX 3200 Systems
17/114