Operating Instructions
5
WARNING! Personal Injury Hazard. Follow the instructions in the documentation
when using the system. The protection provided by the equipment might be
impaired if the equipment is used in a manner not specified by SCIEX.
Start Up the System
WARNING! Electrical Shock Hazard. Make sure that the system can be
disconnected from the mains supply outlet in an emergency. Do not block
the mains supply outlet.
Note:
Before operating the instrument, read the safety information in the section:
.
Prerequisites
• The site requirements specified in the
Site Planning Guide
are met. The
Site Planning
Guide
includes information on the mains supply and connections, compressed air, nitrogen,
roughing pump, ventilation, exhaust, and site clearance requirements. Contact SCIEX for
a copy of the
Site Planning Guide
, if required. For contact information, go to
.
• The source exhaust gas, compressed air, and nitrogen gases are connected to the mass
spectrometer.
• The 4 L source exhaust drain bottle is connected to the exhaust waste connection on the
back of the mass spectrometer and to the laboratory ventilation system.
• The source exhaust hoses are securely clamped at the mass spectrometer, source exhaust
drain bottle, and ventilation connections.
• The mass spectrometer convenience switch is turned off and the mains supply cable is
connected to the mass spectrometer.
• The mass spectrometer and roughing pump mains supply cables are connected to the
200 VAC to 240 VAC mains supply.
• The Ethernet cable is connected to both the mass spectrometer and the computer.
1. Turn on the roughing pump.
The On/Off switch is beside the mains supply input connection on the roughing pump.
Operating Instructions
System User Guide
RUO-IDV-05-6475-C
SCIEX 3200 Systems
47/114