STATIM
900J Operator’s Manual
9
4. Use the cycle buttons to select and change the selected field’s value. To
increase a field’s value, press the Solid Instrument cycle button. Hold the
button down to increase the value.
5. To decrease the value, press the Hollow or Wrapped Instrument cycle
button.
6. To select the next field, press the RUBBER AND PLASTICS cycle button.
7. To save changes and return to the regular operating mode, press the
MULTIFUNCTION button.
8. To quit without making changes, power the Statim OFF.
3.8 Preparing Unit for Use
Once the unit is installed and before any instruments are sterilized, run two
Hollow or Wrapped instrument cycles (See section 4.5 Running a Cycle). Clean
the drawer and the instrument basket using a soft cloth to wipe the inside
surfaces and rinse thoroughly with tap water. Once the instrument basket and
sterilization chamber are clean and dry, coat the interior surfaces with Stat-Dri.
3.9 Shipping the Statim 900J
If you must ship the Statim 900J, follow these instructions:
1. Drain the water from the reservoir. Carefully turn the Statim 900J backwards.
Locate the drain tube beside the power switch on the back of the unit.
Release the drain tube from the clip.
2. Remove the stopper from the end of the drain tube and allow for the water
to drain freely from the reservoir. When the water no longer drips from the
drain tube replace the stopper and secure the drain tube under the clip.