STATIM
900J Operator’s Manual
18
4.6 Stopping a Cycle
To stop a cycle, press the MULTIFUNCTION button. If the MULTIFUNCTION
button is pushed, or the unit detects a problem while operating, the cycle will
stop and the amber active light will flash. Once a cycle has been interrupted, the
MULTIFUNCTION button must be pressed before another cycle can be started.
The display reads the following message:
If the display shows the message, CYCLE FAULT or NOT STERILE, the chamber
contents are not sterile! See section 6. Troubleshooting for more information.
4. Instructions for Use
BUTTON PRESSED
PLEASE WAIT
Содержание STATIM 900J
Страница 31: ...STATIM 900J Operator s Manual 31...