3.3 Connecting the waste bottle
To connect the waste bottle (1), follow these steps:
1. Insert one end of the
Teflon™ exhaust tube
(2) into the exhaust
fitting (3) on the back of
the Statim 900J. Push
the tube into the fitting
as far as it will go to
ensure that the tube is
connected tightly.
2. Cut the tube to length
and slide the nut, ring
and cone onto the tube.
Place the free end of
the tube into the fitting
in the lid of the waste
bottle and hand tighten.
Do not coil the exhaust
tube.
3. Place the waste bottle
near the Statim 900J.
We recommend that
you keep it in a cabinet
below the unit. For convenience, the tube can be routed through a hole, 7
mm (
5
/
16
") in diameter, drilled through the counter-top and secured with the
nylon clamps provided.
4. Unscrew the lid with the copper condenser coil assembly from the waste
bottle. The lid and coil should come out together. Fill the waste bottle with
water to the “MIN” line and replace the lid and copper condenser assembly.
5. If required, a low-level disinfectant prepared according to instructions on
the disinfectant packaging may be added to the waste bottle. This may help
prevent unpleasant odours or discoloration of the waste water if it is not
changed regularly.
3.4 Filling the Statim Reservoir
To fill the reservoir, follow these steps:
1. Remove the reservoir cap from the top of
the Statim 900J and pour only steam-
process distilled water into the reservoir
until nearly full (a maximum of 3 litres).
Never use tap water. We recommend using
a funnel when filling the reservoir to avoid
spills. Pour slowly when filling the top portion
of the reservoir. Use only steam processed
distilled water containing less than 5 ppm
total dissolved solids (having conductivity of
less than 10 μS / cm).
1.
waste Bottle
2.
Teflon™ exhaust tube
3.
exhaust fitting
4. nut
5. ring
6. cone
7.
threaded nut
3
2
4
5
6
7
1
Figure 4
Figure 5
3. Installation