SciCan STATIM 900J Скачать руководство пользователя страница 8

STATIM

 900J Operator’s Manual

8

2.    Replace the reservoir cap to ensure that nothing other than steam-process 

distilled water enters the reservoir. 

3.  Empty the waste bottle and refill to the MIN indicator.

3.5 Priming the Statim 900J pump

To Prime the Statim 900J pump follow  

these steps:
1.    Carefully turn the Statim 900J sideways,  

and locate the drain tube beside the  

power switch, on the back of the unit.  

Gently pull the tube out as far as possible  

so the free end can be positioned over a  

water container.

2.    Remove the stopper from the end of the  

drain tube and allow water to drain from  

the tube into a container for at least  

30 seconds.  

3.    Replace the stopper, and feed the tube back into the unit until secure. 
4.    Plug the power cord into the power cord receptacle on the back of the unit.
5.   Plug the power cord into a standard, grounded, wall outlet.

3.6 Language Selection

To set the language feature for the Statim 900J, follow these steps. Watch the 

blinking cursor on the LCD.
1.  Turn the Statim 900J power switch OFF
2.   Press and hold down the Hollow or Wrapped Instrument Cycle button on 

the front of  the Statim 900J, and turn the power switch at the back of the 

unit ON. Release the Solid Instrument Cycle button. The current language 

selection will appear.

3.   To scroll to the next language press the Hollow or Wrapped Cycle button.  

To scroll to the previous language press the Solid Instrument Cycle button. 

4.   When the desired language feature is indicated on the LCD, press  

the MULTIFUNCTION button shown to save the change and return to  

normal operation. 

3.7 Setting the Time and Date

To set the time and date, follow these steps and watch the blinking cursor  

on the LCD:
1.   Power the Statim OFF.
2.   Press and hold the Solid Instrument cycle button.
3.   While pressing the Solid Instrument cycle button, power the Statim ON.  

The LCD display reads: 

Figure 6

3. Installation

14:23 06/22/2007

HH:MM DD/MM/YYYY

Содержание STATIM 900J

Страница 1: ...Operator s Manual STATIM 900J AUTOCLAVE 95 110106 JP EN R2 Copyright 2017 SciCan Ltd All rights reserved...

Страница 2: ...e Maintenance Schedules 5 4 Spare Parts List 6 Troubleshooting Problems 22 7 The Communication Port 24 7 1The RS232 Communication Port 7 2 Installing the SciCan Data Logger 7 3 Cycle Printout Overview...

Страница 3: ...instruments immediately after they are comfortable to the touch The Statim 900J has three sterilization cycles The Solid Instrument Cycle is used to sterilize solid metal instruments The sterilization...

Страница 4: ...ll not be liable for incidental special or consequential damages caused by any maintenance or services performed on the Statim by a third party or for the use of equipment or parts manufactured by a t...

Страница 5: ...to the operator A situation which may lead to a mechanical failure The following symbols appear in the margins of this book 2 4 1 3 10 8 13 9 Figure 2 5 Tube Mounting Hardware Waste Bottle part numbe...

Страница 6: ...ed surface Electromagnetic Environment The Statim has been tested and meets applicable standards for electromagnetic emissions While the unit does not emit any radiation it may itself be affected by o...

Страница 7: ...le The lid and coil should come out together Fill the waste bottle with water to the MIN line and replace the lid and copper condenser assembly 5 If required a low level disinfectant prepared accordin...

Страница 8: ...3 6 Language Selection To set the language feature for the Statim 900J follow these steps Watch the blinking cursor on the LCD 1 Turn the Statim 900J power switch OFF 2 Press and hold down the Hollow...

Страница 9: ...ow or Wrapped instrument cycles See section 4 5 Running a Cycle Clean the drawer and the instrument basket using a soft cloth to wipe the inside surfaces and rinse thoroughly with tap water Once the i...

Страница 10: ...automatically To Remove The Sterilization Chamber Drawer To remove the sterilization chamber drawer follow these steps 1 Open the drawer by pressing the MULTIFUNCTION button on the keypad 2 Turn the...

Страница 11: ...t be protected from re contamination during transport to the point of use The rapid timing of the Solid Instrument cycle about 6 minutes is sufficient to sterilize instruments between patients Hollow...

Страница 12: ...le load aceptically Best Practice Allow instruments wrapped or unwrapped to dry completely prior to handling Wrapped or pouched instruments must not touch each other to promote drying and enable effec...

Страница 13: ...he correct indicator type was chosen 3 Ensure the chamber was not overloaded Consult the earlier portion of this section for proper loading instructions 4 If the results do not change do not process a...

Страница 14: ...ization cycles each designed to sterilize a specific type of instrument The instruments will remain sterile after a successful cycle until the drawer is opened The types of instruments sterilization r...

Страница 15: ...t Cycle is used to sterilize instruments constructed of the materials listed under the heading Rubber or Plastic Instruments in Preparing and Loading Instruments The maximum recommended load for the R...

Страница 16: ...ns Messages change throughout the cycle For example if the SOLID INSTRUMENT CYCLE is selected the following LCD messages would be displayed if the drawer is ajar STATIM 900J R1 11 SELECT A CYCLE PRESS...

Страница 17: ...ext cycle This ensures that the CYCLE COMPLETE message is not displayed when a new load of unsterilized instruments is placed in the chamber After 30 minutes the LCD message will automatically change...

Страница 18: ...ill stop and the amber active light will flash Once a cycle has been interrupted the MULTIFUNCTION button must be pressed before another cycle can be started The display reads the following message If...

Страница 19: ...eading The water in contact with the hot chamber surfaces evaporates more efficiently Spotting is minimized and instruments dry faster Stat Dri is available from your dealer or from SciCan Cleaning th...

Страница 20: ...rain Daily Waste Bottle Empty the waste bottle every time you refill the reservoir Fill with tap water up to MIN line You can also add some chlorine free disinfectant technician Once Chamber Check the...

Страница 21: ...65S Condenser Bottle 01 101582S Replacement Cap 01 103636S Wire Basket Complete 01 104343S Plug Drain Tubing 01 101562S Statim 900J Drawer 2OZPLUS Stat Dri Plus 2oz bottle w sprayer 32OZPLUS Stat Dri...

Страница 22: ...n filling your Statim 900J Make sure the plug in the drain tube is secured If the leak persists unload the chamber unplug the unit and call your dealer Cycle aborted message NOT STERILE and CYCLE FAUL...

Страница 23: ...in the reservoir and refill with steam process distilled water To drain the unit use the drain tube at the back of the machine Refer to the steps described in Shipping the Statim 900J Message REFILL R...

Страница 24: ...haracter Epson TM U220D C31C515603 CR LF 9600 248 0xF8 Citizen IDP 3110 40 RF 120B CR 9600 N A Star Micro SP212FD42 120 CR 9600 210 0xd2 Star Micro SP216FD41 120 CR LF 9600 210 0xd2 Star Micro SP512MD...

Страница 25: ...plete installation steps STEP 1 Selecting the USB Flash MSD Option 1 Power up unit while pressing the MULTIFUNCTION button to access the User Setup menu 2 Using the Solid Instrument or Hollow or Wrapp...

Страница 26: ...r user char menu using the Solid Instrument cycle button to increase the displayed value by one and the Hollow or Wrapped Instrument Cycle button to increase that value by ten enter the value 32 0x20...

Страница 27: ...low or Wrapped Instrument Cycle buttons scroll to the Unit ID Setup and press the Rubber and Plastic Cycle button to select it 3 From the Setup Unit ID menu use the Solid Instrument or Hollow or Wrapp...

Страница 28: ...art Time of Sterilization 3 14 start of D phase 11 Temp Press Time at start of sterilization D phase 12 17 Cycle pararmeters at 30 second intervals 18 Temp Press Time of end of sterilization phase end...

Страница 29: ...ice from the authorized SciCan Ltd dealer and such invoice identifies the item by serial number and clearly states the date of purchase No other validation is acceptable After one year all SciCan Ltd...

Страница 30: ...Height 282 mm 11 1 inches Chamber Size Length 190 mm 7 5 inches Width 90 mm 3 54 inches Height 53 mm 2 1 inches Reservoir Capacity Usable volume 3 0 litres 102 fl oz Weight Without water 13 Kg 29 lbs...

Страница 31: ...STATIM 900J Operator s Manual 31...

Отзывы: