After venting is complete, the drawer will open automatically and the instruments
may be removed. The display reads:
Be careful. The instruments and metal parts of the drawer will be hot. Hot steam
will escape as the drawer opens.
This message is displayed for 30 minutes as a reminder of a successfully
completed cycle. When sterilized instruments are removed press the
MULTIFUNCTION button to clear the CYCLE COMPLETE message, and prepare
the machine for the next cycle. This ensures that the CYCLE COMPLETE
message is not displayed when a new load of unsterilized instruments is placed
in the chamber.
After 30 minutes the LCD message will automatically change to:
Once exposed to a non-sterile environment at the end of a complete cycle, the
continued sterility of the chamber contents can no longer be assured. Therefore,
the Statim 900J should only be used for the prevention of cross-contamination,
unless other steps are taken to ensure continued sterility.
SELECT A CYCLE
PRESS TO CLOSE
4. Instructions for Use
SOLID INSTRUMENTS
PRESSURIZING 130ºC
STERILIZING
134ºC 212 kPa 3:15
SOLID INSTRUMENTS
VENTING 115ºC
SOLID INSTRUMENTS
CYCLE COMPLETE 6:02