For detailed instructions on how to handle, use and dispose of both the
biological and chemical indicators, please consult the product literature
accompanying the 3M Attest™ biological indicators or contact the manufacturer
directly.
To use the indicators with the Statim 900J, follow these steps:
1. Place the appropriate biological indicator in the Statim 900J chamber.
2. Process the load in the sterilizer according to your usual practice.
3. Ensure that the message “Cycle Complete” is displayed on the LCD after
the cycle is finished.
4. Recover the biological and / or chemical indicator and process further
according to the literature that accompanied the indicator.
At the first indication of a potential sterilization failure:
1. Do not process any more instruments until favourable test results have
been returned.
2. Ensure the correct indicator type was chosen.
3. Ensure the chamber was not overloaded. Consult the earlier portion of this
section for proper loading instructions.
4. If the results do not change, do not process any more instruments within the
Statim 900J and contact your SciCan dealer for further assistance.
Because the turnaround time for the 3M Attest™ is up to 48 hours, it is
recommended that the tests be conducted so that the incubation period occurs
during a period of planned downtime such a the last cycle before a weekend.
4.3 Instrument Mass Guide
Instrument
Typical Instrument Mass
Scissors
30 g (0.96 ozt)
Dental Scalers
20 g (0.64 ozt)
Forceps
15 g (0.48 ozt)
Dental handpiece
40 to 60 g (1.3 to 1.9 ozt)
Unwrapped instrument rack
225 g (7.23 ozt)
Suction cannula
10 g (0.32 ozt)
Plastic mouth mirror
8 g (0.3 ozt)
Impression Tray
15 to 45 g (0.48 to 1.5 ozt)
Plastic x-ray positioning ring
20 g (0.64 ozt)
NOTE
: The above weights are to be used as reference only. For exact weights
of your instruments, consult the manufacturer's specifications.
4. Instructions for Use