5.1 Cleaning the Chamber
It is good clinical practice to keep your Statim 900J sterilization chamber clean.
We recommend that the interior surface of the chamber be cleaned at least once
a week with dishwashing soap or mild detergent that does not contain chlorine.
Remove the drawer from the unit to make cleaning easier. See, section 4.1.
Afterwards, rinse thoroughly to remove all traces of soap. After the chamber is
clean treat the entire inside surface with Stat-Dri. Stat-Dri causes water to form
an even coat on treated surfaces, without beading. The water in contact with
the hot chamber surfaces evaporates more efficiently. Spotting is minimized and
instruments dry faster. Stat-Dri is available from your dealer or from SciCan.
Cleaning the inside of the chamber is very important if you regularly sterilize
lubricated instruments.
5.2 Cleaning the Exterior Surfaces
Use a soft cloth moistened with a mild cleaning solution or mild disinfectant to
clean all outer surfaces. Do not use solvents or harsh chemicals.
5.3 Preventative Maintenance Schedules
To ensure trouble-free performance, both the operator and the dealer must
follow a preventative maintenance schedule.
NOTE
: Please refer to your National, Regional, State or Safety laws for any
additional reoccurring user testing that may be required.