STATIM

 900J Operator’s Manual

14

4.4 Selecting a Cycle

The Statim 900J Autoclave has been designed expressly for the rapid 

sterilization of instruments intended for immediate use. This allows you to 

sterilize instruments between  patients. Quick drying and cooling of the load 

at the end of the cycle is achieved through evaporation after the automatic 

opening of the sterilzation chamber drawer. Because of the unique operation 

of the Statim 900J, once the instruments are exposed to ambient conditions, 

their continued sterility cannot be assured. It is recommended to use these 

instruments immediately after they are comfortable to the touch. Each cycle can 

be selected by pressing the Solid Instrument, Hollow or Wrapped Instrument, or 

Rubber / Plastic buttons.
The types of instruments, sterilization requirements and a graph depicting each 

cycle characteristics are described over the following pages.
The Statim 900J has three sterilization cycles, each designed to sterilize a 

specific type of instrument. The instruments will remain sterile after a successful 

cycle until the drawer is opened. The types of instruments, sterilization 

requirements, and a graph depicting each cycle are described over the next few 

pages. Consult the Instrument Mass Guide in Section 4.3 for information on how 

to make up an appropriate load for the masses specified for individual cycles.

1. Solid Instrument Cycle
The Solid Instrument Cycle is used to sterilize solid metal instruments free from 

deep cavities or holes, such as pliers, burrs, scalers and  forceps.
The maximum recommended load for the Solid Instrument Cycle is 1.0 kg.  
To select this cycle, press the Solid Instrument Cycle button. The cycle then 

starts automatically. This cycle is illustrated in the graph below. During the 

sterilizing phase of this cycle, the temperature in the chamber is maintained at 

a minimun of 134ºC for 3.5 minutes. See Preparing and Loading Instruments 

before running this cycle.

SOLID 134°C / 3.5 min

A

B

D

E

C

Warming up

Conditioning

Pressurizing

Sterilizing

Venting

4. Instructions for Use

Содержание STATIM 900J

Страница 1: ...Operator s Manual STATIM 900J AUTOCLAVE 95 110106 JP EN R2 Copyright 2017 SciCan Ltd All rights reserved...

Страница 2: ...e Maintenance Schedules 5 4 Spare Parts List 6 Troubleshooting Problems 22 7 The Communication Port 24 7 1The RS232 Communication Port 7 2 Installing the SciCan Data Logger 7 3 Cycle Printout Overview...

Страница 3: ...instruments immediately after they are comfortable to the touch The Statim 900J has three sterilization cycles The Solid Instrument Cycle is used to sterilize solid metal instruments The sterilization...

Страница 4: ...ll not be liable for incidental special or consequential damages caused by any maintenance or services performed on the Statim by a third party or for the use of equipment or parts manufactured by a t...

Страница 5: ...to the operator A situation which may lead to a mechanical failure The following symbols appear in the margins of this book 2 4 1 3 10 8 13 9 Figure 2 5 Tube Mounting Hardware Waste Bottle part numbe...

Страница 6: ...ed surface Electromagnetic Environment The Statim has been tested and meets applicable standards for electromagnetic emissions While the unit does not emit any radiation it may itself be affected by o...

Страница 7: ...le The lid and coil should come out together Fill the waste bottle with water to the MIN line and replace the lid and copper condenser assembly 5 If required a low level disinfectant prepared accordin...

Страница 8: ...3 6 Language Selection To set the language feature for the Statim 900J follow these steps Watch the blinking cursor on the LCD 1 Turn the Statim 900J power switch OFF 2 Press and hold down the Hollow...

Страница 9: ...ow or Wrapped instrument cycles See section 4 5 Running a Cycle Clean the drawer and the instrument basket using a soft cloth to wipe the inside surfaces and rinse thoroughly with tap water Once the i...

Страница 10: ...automatically To Remove The Sterilization Chamber Drawer To remove the sterilization chamber drawer follow these steps 1 Open the drawer by pressing the MULTIFUNCTION button on the keypad 2 Turn the...

Страница 11: ...t be protected from re contamination during transport to the point of use The rapid timing of the Solid Instrument cycle about 6 minutes is sufficient to sterilize instruments between patients Hollow...

Страница 12: ...le load aceptically Best Practice Allow instruments wrapped or unwrapped to dry completely prior to handling Wrapped or pouched instruments must not touch each other to promote drying and enable effec...

Страница 13: ...he correct indicator type was chosen 3 Ensure the chamber was not overloaded Consult the earlier portion of this section for proper loading instructions 4 If the results do not change do not process a...

Страница 14: ...ization cycles each designed to sterilize a specific type of instrument The instruments will remain sterile after a successful cycle until the drawer is opened The types of instruments sterilization r...

Страница 15: ...t Cycle is used to sterilize instruments constructed of the materials listed under the heading Rubber or Plastic Instruments in Preparing and Loading Instruments The maximum recommended load for the R...

Страница 16: ...ns Messages change throughout the cycle For example if the SOLID INSTRUMENT CYCLE is selected the following LCD messages would be displayed if the drawer is ajar STATIM 900J R1 11 SELECT A CYCLE PRESS...

Страница 17: ...ext cycle This ensures that the CYCLE COMPLETE message is not displayed when a new load of unsterilized instruments is placed in the chamber After 30 minutes the LCD message will automatically change...

Страница 18: ...ill stop and the amber active light will flash Once a cycle has been interrupted the MULTIFUNCTION button must be pressed before another cycle can be started The display reads the following message If...

Страница 19: ...eading The water in contact with the hot chamber surfaces evaporates more efficiently Spotting is minimized and instruments dry faster Stat Dri is available from your dealer or from SciCan Cleaning th...

Страница 20: ...rain Daily Waste Bottle Empty the waste bottle every time you refill the reservoir Fill with tap water up to MIN line You can also add some chlorine free disinfectant technician Once Chamber Check the...

Страница 21: ...65S Condenser Bottle 01 101582S Replacement Cap 01 103636S Wire Basket Complete 01 104343S Plug Drain Tubing 01 101562S Statim 900J Drawer 2OZPLUS Stat Dri Plus 2oz bottle w sprayer 32OZPLUS Stat Dri...

Страница 22: ...n filling your Statim 900J Make sure the plug in the drain tube is secured If the leak persists unload the chamber unplug the unit and call your dealer Cycle aborted message NOT STERILE and CYCLE FAUL...

Страница 23: ...in the reservoir and refill with steam process distilled water To drain the unit use the drain tube at the back of the machine Refer to the steps described in Shipping the Statim 900J Message REFILL R...

Страница 24: ...haracter Epson TM U220D C31C515603 CR LF 9600 248 0xF8 Citizen IDP 3110 40 RF 120B CR 9600 N A Star Micro SP212FD42 120 CR 9600 210 0xd2 Star Micro SP216FD41 120 CR LF 9600 210 0xd2 Star Micro SP512MD...

Страница 25: ...plete installation steps STEP 1 Selecting the USB Flash MSD Option 1 Power up unit while pressing the MULTIFUNCTION button to access the User Setup menu 2 Using the Solid Instrument or Hollow or Wrapp...

Страница 26: ...r user char menu using the Solid Instrument cycle button to increase the displayed value by one and the Hollow or Wrapped Instrument Cycle button to increase that value by ten enter the value 32 0x20...

Страница 27: ...low or Wrapped Instrument Cycle buttons scroll to the Unit ID Setup and press the Rubber and Plastic Cycle button to select it 3 From the Setup Unit ID menu use the Solid Instrument or Hollow or Wrapp...

Страница 28: ...art Time of Sterilization 3 14 start of D phase 11 Temp Press Time at start of sterilization D phase 12 17 Cycle pararmeters at 30 second intervals 18 Temp Press Time of end of sterilization phase end...

Страница 29: ...ice from the authorized SciCan Ltd dealer and such invoice identifies the item by serial number and clearly states the date of purchase No other validation is acceptable After one year all SciCan Ltd...

Страница 30: ...Height 282 mm 11 1 inches Chamber Size Length 190 mm 7 5 inches Width 90 mm 3 54 inches Height 53 mm 2 1 inches Reservoir Capacity Usable volume 3 0 litres 102 fl oz Weight Without water 13 Kg 29 lbs...

Страница 31: ...STATIM 900J Operator s Manual 31...

Отзывы: