93
CF #
Description
Possible causes
Phase of the cycle
CF 83
Bottom band heater
temperature sensor
failure
Bottom band heater temperature
sensor failure
Controller board failure
CF 84
Venturi water reservoir
temperature sensor
failure
Venturi water reservoir temperature
sensor failure
Controller board failure
CF 90
Corrupted/not
initialized chamber
thermocouple
calibration
Corrupted/not initialized chamber
thermocouple calibration
CF 91
Corrupted/not
initialized pressure
calibration
Corrupted/not initialized pressure
calibration
CF 92
Controller board
temperature high
Air filter clogged, poor air ventilation
CF 95
Communication Lost
Harness cable between the User
interface board and controller board
damaged or disconnected
CF 98
Communication with the
ADC failed
Controller board damaged
Содержание STATCLAVE G4
Страница 5: ...5 Diagram of key systems and components...
Страница 6: ...6...
Страница 17: ...17 Checking package contents...
Страница 18: ...18 Positioning a unit Unit dimensions and operating environment...
Страница 20: ...20 Direct to drain connection recommended...
Страница 21: ...21 Connecting to a waste bottle...
Страница 22: ...22 Connecting to the Internet Connecting to a wired network Connecting to a wireless network...
Страница 23: ...23...
Страница 25: ...25 Connecting an auxiliary pump...
Страница 28: ...28...
Страница 30: ...30 Starting and stopping a cycle...
Страница 50: ...50 Registering for online access...
Страница 51: ...51 First start up...
Страница 52: ...52...
Страница 58: ...58 Draining the unit for cleaning and shipping...
Страница 61: ...61 Replacing the bacteriological filter Replacing the door seal...
Страница 75: ...75...
Страница 107: ...107 APPENDIX A STATCLAVE G4 plumbing diagram...
Страница 108: ...108 APPENDIX B STATCLAVE G4 electrical schematic...