84
PROBLEM
POSSIBLE CAUSE
POSSIBLE SOLUTION
The air intake on the frame
and/or the cover is
obstructed, or the heat
exchanger is not sufficiently
ventilated.
Remove all possible obstructions from
the air intake and heat exchanger.
Check that the device is not in direct
contact with walls or surfaces (see
Section 2 Location and Installation).
Vacuum test has failed
The unit chamber was hot
when the test was initiated.
Ensure stand-by chamber warming
system is turned OFF. (See Section 7.4
Setting the Stand-by Mode). Attempt a
second Vacuum Test making sure the
chamber is cooled to room
temperature.
Water is pooling under the unit There is a leak in the system. Contact your SciCan dealer.
Bowie-Dick test has failed
Attempt a second Bowie-Dick test. If it
fails, contact your SciCan dealer.
No cycles are stored in the
unit’s memory
Logic board configuration
issue
Check unit serial number to see if it
was accurately updated after a logic
board service. If number consists of
zeros, call SciCan.
Содержание STATCLAVE G4
Страница 5: ...5 Diagram of key systems and components...
Страница 6: ...6...
Страница 17: ...17 Checking package contents...
Страница 18: ...18 Positioning a unit Unit dimensions and operating environment...
Страница 20: ...20 Direct to drain connection recommended...
Страница 21: ...21 Connecting to a waste bottle...
Страница 22: ...22 Connecting to the Internet Connecting to a wired network Connecting to a wireless network...
Страница 23: ...23...
Страница 25: ...25 Connecting an auxiliary pump...
Страница 28: ...28...
Страница 30: ...30 Starting and stopping a cycle...
Страница 50: ...50 Registering for online access...
Страница 51: ...51 First start up...
Страница 52: ...52...
Страница 58: ...58 Draining the unit for cleaning and shipping...
Страница 61: ...61 Replacing the bacteriological filter Replacing the door seal...
Страница 75: ...75...
Страница 107: ...107 APPENDIX A STATCLAVE G4 plumbing diagram...
Страница 108: ...108 APPENDIX B STATCLAVE G4 electrical schematic...