3. FUNCTIONS AND CYCLES
The STATCLAVE G4 is a dynamic air removal (pre-vacuum) table-top steam sterilizer that uses steam to
sterilize wrapped and unwrapped instrument loads typically used in dental and medical offices. It has six
validated sterilization cycles with optimized drying for fast, effective instrument processing. An additional
custom cycle can be configured using one of three temperature settings but this cycle must be validated by
the user.
Touchscreen overview
The home screen default setting shows icons for the four most common cycles: Solid Unwrapped, Solid
Wrapped, Hollow Unwrapped and Hollow Wrapped. The “…” icon provides access to additional cycles. These
include: Rubber & Plastic, Textiles/Porous and a Custom Cycle. The Test Cycles icon provides access to a
Bowie Dick Test and a Vacuum Test.
Содержание STATCLAVE G4
Страница 5: ...5 Diagram of key systems and components...
Страница 6: ...6...
Страница 17: ...17 Checking package contents...
Страница 18: ...18 Positioning a unit Unit dimensions and operating environment...
Страница 20: ...20 Direct to drain connection recommended...
Страница 21: ...21 Connecting to a waste bottle...
Страница 22: ...22 Connecting to the Internet Connecting to a wired network Connecting to a wireless network...
Страница 23: ...23...
Страница 25: ...25 Connecting an auxiliary pump...
Страница 28: ...28...
Страница 30: ...30 Starting and stopping a cycle...
Страница 50: ...50 Registering for online access...
Страница 51: ...51 First start up...
Страница 52: ...52...
Страница 58: ...58 Draining the unit for cleaning and shipping...
Страница 61: ...61 Replacing the bacteriological filter Replacing the door seal...
Страница 75: ...75...
Страница 107: ...107 APPENDIX A STATCLAVE G4 plumbing diagram...
Страница 108: ...108 APPENDIX B STATCLAVE G4 electrical schematic...