4. MAINTENANCE
General information
Regular maintenance will ensure the safe and efficient operation of your STATCLAVE G4. Before conducting
any of the cleaning and maintenance procedures described in this chapter, power OFF the unit and
disconnect it from its power source.
WARNING: HOT SURFACES.
The STATCLAVE G4 chamber’s stand-by mode maintains an optimal
operating temperature during working hours. Unless this feature is disabled, the chamber will remain hot
between cycles throughout the work day. Take care to ensure the STATCLAVE G4 is properly cooled before
accessing the chamber to perform any maintenance.
DO
always use SciCan replacement parts.
DON’T
use abrasive cloths, metal brushes or metal-cleaning products, whether solids or liquids, to clean the
device or sterilization chamber.
Maintenance message
WHEN A MAINTENANCE MESSAGE APPEARS, the user has 2 options:
1.
Pressing OK to acknowledge the message. The user can then continue to use the STATCLAVE G4 or
perform the required maintenance. When the user presses OK, the maintenance notification counter
will restart the counter, regardless of whether or not the user has performed the maintenance.
2.
Pressing REMIND LATER will prompt the STATCLAVE G4 to repeat the message 24 hours later.
Preventative scheduled maintenance
OPERATOR
DAILY
Wipe the door gasket clean with a damp, lint-free
cloth.
Clean external surfaces with a damp, lint-free
cloth.
WEEKLY
Clean the chamber and, if applicable, the waste
water bottle.
Disinfect external surfaces.
Содержание STATCLAVE G4
Страница 5: ...5 Diagram of key systems and components...
Страница 6: ...6...
Страница 17: ...17 Checking package contents...
Страница 18: ...18 Positioning a unit Unit dimensions and operating environment...
Страница 20: ...20 Direct to drain connection recommended...
Страница 21: ...21 Connecting to a waste bottle...
Страница 22: ...22 Connecting to the Internet Connecting to a wired network Connecting to a wireless network...
Страница 23: ...23...
Страница 25: ...25 Connecting an auxiliary pump...
Страница 28: ...28...
Страница 30: ...30 Starting and stopping a cycle...
Страница 50: ...50 Registering for online access...
Страница 51: ...51 First start up...
Страница 52: ...52...
Страница 58: ...58 Draining the unit for cleaning and shipping...
Страница 61: ...61 Replacing the bacteriological filter Replacing the door seal...
Страница 75: ...75...
Страница 107: ...107 APPENDIX A STATCLAVE G4 plumbing diagram...
Страница 108: ...108 APPENDIX B STATCLAVE G4 electrical schematic...