43
Using the service menu
To access the service menu, select the SETTINGS icon from the home screen, press on the image of the
technician and enter the service code 7919 on the keypad.
Menu item
Submenu item
What to do with it
PRODUCTION TOOLS
Calibration
To calibrate the unit using a
probe, set calibration to ON so
that you can run a cycle and
adjust the temperature offset.
Burn-in mode
Conductivity Setup
DIAGNOSTIC TOOLS
This menu contains a number of tools technicians can use to more accurately diagnose problems.
Device Status/Test
Repeater Mode
Use to run a series of cycles in
succession. Can be set to repeat
1 to 5 cycles in a row or set it to
ON to run the same cycle over
again until you stop it.
Repeater count
Debug Mode
To display the debug screen
while running a cycle, set Debug
Screen to ON.
Clear CF Printouts
Calibration Offsets
To calibrate the unit using a
probe, set calibration to ON so
that you can run a cycle and
adjust the temperature offset.
FACTORY DEFAULT
Cycle Number
Содержание STATCLAVE G4
Страница 5: ...5 Diagram of key systems and components...
Страница 6: ...6...
Страница 17: ...17 Checking package contents...
Страница 18: ...18 Positioning a unit Unit dimensions and operating environment...
Страница 20: ...20 Direct to drain connection recommended...
Страница 21: ...21 Connecting to a waste bottle...
Страница 22: ...22 Connecting to the Internet Connecting to a wired network Connecting to a wireless network...
Страница 23: ...23...
Страница 25: ...25 Connecting an auxiliary pump...
Страница 28: ...28...
Страница 30: ...30 Starting and stopping a cycle...
Страница 50: ...50 Registering for online access...
Страница 51: ...51 First start up...
Страница 52: ...52...
Страница 58: ...58 Draining the unit for cleaning and shipping...
Страница 61: ...61 Replacing the bacteriological filter Replacing the door seal...
Страница 75: ...75...
Страница 107: ...107 APPENDIX A STATCLAVE G4 plumbing diagram...
Страница 108: ...108 APPENDIX B STATCLAVE G4 electrical schematic...