Operating manual 810378-BAL-EN-170330
Mach/30 2017
Page 49 of 54
11.2.4.
I/O data definition (area, unit, standard)
Data
Unit
Min
Max smart-
E 310
Max smart-E
510
Stan
dard
Remarks
U target
kV
10
100
0
kV target val-
ue/limiting value
U min
kV
10
100
0
Limit value for Umin-
switch off
I Limit*
uA
10
300
800 / 600
0-80 kV
– 800
81-100 kV - 600
0
Target value for cur-
rent limiting value
(800 below 80 kV)
I max.*
uA
10
300
800
0
Limit value for Imax -
switch off
U actual
100
V
0
1095
1240
-
Actual value kV
I actual
uA
0
667
1130
-
Actual current
kV-
Rampe
kV/s
10
100
0
Speed when starting
up
voltage
Sensitivity
*
of the electrical
arc detection
0
10
0
0: Electrical arc de-
tection off
1~10: Highest to
lowest
sensitivity,
Typically this parameter can be set to 5.
Do not switch off electrical arc detection in normal operation!
Values outside of these areas (min, max) are ignored. The warning "invalid parame-
ters" is output.
Caution: The Byte sequence is Little-Endian. The Byte sequence can be taken from the
following examples:
Example: From a smart-E 510 received data:
Control word
U Target value
U min
I Limit
Imax
HV ON
20 kV
5 kV
256 uA
50 uA
Byte 0 Byte 1 Byte 2 Byte 3 Byte 4 Byte 5 Byte 6 Byte 7 Byte 8 Byte 9
0x01
0x00
0x14
0x00
0x05
0x00
0x00
0x01
0x32
0x00
kV ramp
Sensitivity*
10 kV/s
5
Byte
10
Byte
11
Byte
12
Byte
13
0x0a
0x00
0x05
0x00
Example: From a smart-E 510 sent data:
Status word
Service code
U actual
I actual
20 kV
10 uA
Byte 0 Byte 1 Byte 2 Byte 3 Byte 4 Byte 5 Byte 6 Byte 7 Byte 8 Byte 9
0xC1
0x00
0x00
0x00
0x00
0x00
0xC8
0x00
0x0A
0x00