SCHILLER AT-1 / AT-1smartprint ECG Unit
SERVICE HANDBOOK Issue b June 2001
Art.No. 2.540022
ECG Amplifier Reference Voltage
• Switch the unit on and measure the voltage difference between the +2 V
reference and the -2 V reference on pins 1 and 7 of operational amplifier U100.
Adjust the trimmer VR2 to achieve a voltage difference of 4000 mV +/- 2 mV.
• Obtain a service printout by pressing
ALT - 0 - 3 - any key
• Ensure that the Uref+ and the Uref- measurements are both 2000 mV +/- 20 mV.
Ensure the the Udiff reading is 4000 mV. Adjust VR3 to achieve a Udiff reading of
4000 mV +/- 20 mV.
• Reassemble the unit as detailed in Chapter 4. Recheck the voltage by again
obtaining a service printout.
VR2
VR3
Oper. amp. U100
Pin 1 low/left