General Functional Check Procedure
The procedure detailed here is a general confidence check of the unit after an internal
module or board has been replaced. It is not a full functional test (which can only be
carried out with dedicated equipment in the factory) but is intended to provide a
general confidence check in all the major AT-1 / AT-1smartprint functional areas. The
instructions given here are guides to the basic functions. If more operating information
is required (general settings, comprehensive menu guides etc.) please refer to Chapter
1 in this publication or the relevant User Manual.
To carry out the general AT-1 / AT-1smartprint functional check procedure, proceed as
follows:
1. Connect mains power to the unit and ensure that the green mains LED lights.
2. Switch the unit on by pressing the
key on the keyboard. Ensure that the LED
lights flash for about a second.
3. Carry out the printer check detailed on the next page.
4. Connect an ECG simulator to the ECG connector on the side panel and switch on.
5. Press the
key and ensure that three leads are printed and are of good
quality.
6. Press the
key and wait approximately 10 seconds for the printout to
commence. Ensure that the printout is accurate and of good quality.
7. Switch the unit off
and leave connected to the mains supply for 10 hours or
more to charge the battery.
8. Disconnect the mains and switch the unit on. Ensure that the Battery LED is lit. Run
the unit on battery power for approximately one hour. Ensure that the battery LED
flashes when the battery has limited capacity (not before 4 hours).