![Schiller CARDIOVIT AT-1 Скачать руководство пользователя страница 60](http://html1.mh-extra.com/html/schiller/cardiovit-at-1/cardiovit-at-1_service-handbook_1212075060.webp)
60
Printer Tray Assembly and Thermal Printer
Prerequisites
• The Warnings and Cautions at the beginning of the Chapter must be observed.
• The top housing must be removed as detailed previously. All external cable
assemblies must be disconnected.
Tools
• Posi-drive screwdriver
• Spring compressor or pliers
Part Numbers
The part numbers for the Thermal Printer Assembly, the printhead and the motor are
given in Chapter 6.
E = Paper mark detector
F = Printer motor
CAUTION
THE THERMAL PRINTER CONTAINS STATIC SENSITIVE COMPONENTS. OBSERVE ANTISTATIC
PRECAUTIONS.
Printer / Paper Tray Assembly Removal
1. Remove the following connectors from the Control and Power Supply Board MK
11-10:
•
data connector to the thermal printer
(A)
•
paper mark connector
(B)
•
printer motor connector
(C)
•
earth connector to the printhead
(D)
2. Unscrew the four retaining screws and remove the complete paper tray / printer
assembly.
Printer/Paper Tray Assembly Replacement
1. Position and secure the assembly in the base with the four retaining screws.
2. Reconnect the following connectors to the Control and Power Supply Board MK 11-10:
•
data connector to the thermal printer
(A)
•
paper mark connector
(B)
•
printer motor connector
(C)
•
earth connector
(D)
3. Replace the Top Housing, see page 57.
A
B
C
D
E
F
Retaining
screws