![Schiller CARDIOVIT AT-1 Скачать руководство пользователя страница 64](http://html1.mh-extra.com/html/schiller/cardiovit-at-1/cardiovit-at-1_service-handbook_1212075064.webp)
64
Battery Pack
The battery pack is held in position with double-sided tape.
Prerequisites
• The Warnings and Cautions at the beginning of the Chapter must be observed.
• The top housing must be removed and all external cable assemblies
disconnected.
Tools and Equipment
Double-sided tape
Parts
The part numbers of all replaceable items are given in Chapter 6.
Battery Pack Removal
WARNING
THE MAINS SUPPLY MUST BE DISCONNECTED DURING THIS PROCEDURE
To remove the Battery Pack proceed as follows:
1. Ensure that the unit is switched off and that the mains is disconnected.
2. Disconnect the two bayonet connectors and remove the Battery Pack by gently
pulling away from the base.
Battery Pack Replacement
1. Stick a piece of double-sided tape to the bottom of the battery pack.
2. Position the battery pack with the connectors towards the inside of the unit as
shown above.
3. Press firmly on the battery, so the double-sided tape sticks to the base assembly.
4. Connect the bayonet connectors for the red and black lead as shown above.
5. Replace the top housing and reassemble the unit as described on page 57.
red (live)
black (neutral)