![Schiller CARDIOVIT AT-1 Скачать руководство пользователя страница 61](http://html1.mh-extra.com/html/schiller/cardiovit-at-1/cardiovit-at-1_service-handbook_1212075061.webp)
61
SCHILLER AT-1 / AT-1smartprint ECG Unit
SERVICE HANDBOOK Issue b June 2001
Art.No. 2.540022
Thermal Printer Removal
1. Remove the top housing and the printer tray assembly as outlined above.
2. Turn the printer / paper tray assembly upside-down and unscrew the two printer
retaining screws.
3. Gently remove the printhead taking care to retain the two tensioning screws.
Thermal Printer Replacement.
To replace the thermal printer proceed as follows:
1. Position the printer in the printer / paper tray assembly so that the printer mounting
plate lips slot into the dedicated cutouts in the assembly; secure with the two
retaining screws. Ensure that the cable assemblies from the printer to the interface
PCB are not caught and are not strained.
2. Using a spring compressor or a pair of pliers, insert the two tensioning springs so
that the springs are positioned over the outer two moulded spring supports and in
the indent (hole) in the printer mounting plate
3. Replace the printer / paper tray assembly as described on page 60.
4. Replace the top housing as described on page 57.
Checks, Tests and Adjustments after Printer Replacement
Check the print quality as described in Chapter 3.
Printer Motor Removal and Replacement
1. Remove the top housing and the printer / paper tray assembly as
outlined above.
2. Turn the printer / paper tray assembly on the side to get access to
the printer motor.
3. Unscrew and remove the two fixing screws, and carefully lift the
motor out of the assembly.
4. Replace the motor, carefully put it in place and fasten the two
retaining screws.
5. Replace the printer / paper tray assembly and the top housing as explained
above.
Содержание CARDIOVIT AT-1
Страница 14: ...14 The Keyboard ...
Страница 44: ...44 ...
Страница 46: ...46 Fault Diagnosis Chart ...
Страница 74: ...74 ...
Страница 78: ...78 ...
Страница 82: ...82 ...