![Schiller CARDIOVIT AT-1 Скачать руководство пользователя страница 13](http://html1.mh-extra.com/html/schiller/cardiovit-at-1/cardiovit-at-1_service-handbook_1212075013.webp)
13
SCHILLER AT-1 / AT-1smartprint ECG Unit
SERVICE HANDBOOK Issue b June 2001
Art.No. 2.540022
Location & Power
Location
Do not keep or operate the device in a wet, moist, or dusty environment. Also avoid
exposure to direct sunlight or heat from other sources. Do not allow the unit to come
into contact with acidic vapours or liquids, as such contact may cause irreparable
damage. The unit should not be placed near X-ray or diathermy units, large transformers
or motors. The unit must be placed on a flat surface and must not be operated in areas
where there is any danger of explosion.
Power Supply
The mains connection is on the rear of the unit. The mains indicator lamp on the keyboard
is always lit when the unit is connected to the mains supply. The unit can either be
operated from the mains supply or from the built-in rechargeable battery. The power
source is indicated by the respective indicator lamp. When battery capacity is limited,
the battery symbol flashes on and off.
To recharge the battery, connect the device to the mains supply by means of the
supplied power cable. A totally discharged battery needs less than 15 hours to be fully
recharged (60% in less than 3 hours, 90% in less than 7 hours). A fully charged battery
gives approximately 2 hours of normal use. The unit can remain connected to the mains
supply without any danger of damage to either the battery or the unit.
Switching On and Off
The AT-1 / AT-1 smartprint is switched on with the green ON key
and off by means of the red OFF key
The unit is automatically switched off after 5 minutes (30 seconds if battery capacity is
limited) if no key is pressed and the patient cable is not connected.
Potential Equalisation
If the A T-1 / AT-1 smartprint is used in conjunction with other patient connected equipment,
we recommend that the potential equalisation stud on the rear of the unit is connected
to the building common ground with the yellow/green ground cable (Part-no. 2.310005).
When working from an emergency vehicle, the vehicle common ground can be used.
WARNING !
THE UNIT MUST BE GROUNDED WHEN THE EXTERNAL PRINTER IS CONNECTED.