71
SCHILLER AT-1 / AT-1smartprint ECG Unit
SERVICE HANDBOOK Issue b June 2001
Art.No. 2.540022
Battery Charge Voltage
Precautions and Requirements
The unit must be placed on an antistatic mat and antistatic precautions observed when
any maintenance is carried out on the AT-1 / AT-1 smartprint. The room temperature
should be between 18 and 28 °C.
Tools and Equipment
• Digital Multimeter
• Small flat bladed screwdriver
• Resistor 2.7 kOhms, 250 mW
Procedure
The battery charge voltage is nominally 13.5 V (with a nominal maximum current of
approximately 5 mA). Adjust and check as follows:
• DISCONNECT THE MAINS SUPPLY !
• Disassemble the unit as detailed in Chapter 4 and remove the ECG interface
board MK 11-2.
• Remove the two battery connectors and connect a 2.7 kOhm 250 mW resistor
across the two connectors (TP12 and TP13) to simulate a discharged battery.
Connect the digital multimeter across the resistor.
• Connect the mains supply.
WARNING
EXERCISE CARE - BEWARE THAT POTENTIALLY LETHAL VOLTAGES ARE PRESENT.
• Adjust VR4 to obtain a charging voltage of 13.5 V.
• Disconnect the mains supply and reassemble the unit as detailed in Chapter 4.
TP 12
TP 13
Battery Charge Voltage VR4