68
Introduction
This chapter provides the procedures necessary to check and adjust all service settings.
Every procedure is self-contained and details the tools required to carry out adjustments,
and the test equipment necessary. Any adjustments, special checks or functional tests
that are required on the module, or on associated modules or software, are also detailed.
In-text illustrations support the text where required and provide location details of
connectors, test points and adjustment potentiometers.
Specific warnings and cautions are given in the text where applicable.
The part numbers for all replaceable modules are given in Chapter 6.
The AT-1 / AT-1smartprint has the following adjustments:
• Battery Charge Voltage (VR4 on the control and power supply board MK 11-10)
• ECG Amplifier reference voltage and ramp slope (VR2 and VR3 on the ECG
interface board MK 11-2)
We recommend that the reference voltages are checked every year.