Care & Maintenance
12 Monthly Check
The unit should undergo a technical safety check every 12 months. This safety check
should include the following:
• Visual inspection of the unit and cables.
• Electrical safety tests according to IEC 601-1 Clause 19.
• Functional tests according to Chapter 3.
The test results must be documented.
Cleaning the Casing
CAUTION
SWITCH THE UNIT OFF BEFORE CLEANING AND DISCONNECT THE MAINS AND
THE EXTERNAL PRINTER (IF PRESENT). DO NOT, UNDER ANY CIRCUMSTANCES,
IMMERSE THE APPARATUS INTO A CLEANING LIQUID OR STERILIZE WITH HOT
WATER, STEAM, OR AIR.
The casing of the AT-1 / AT-1smartprint can be cleaned with a soft damp cloth on the
surface only. Where necessary a domestic non-caustic cleaner can be used for grease
and finger marks.
Patient Cable
CAUTION
ALIGN THE LEADS IN SUCH A WAY AS TO PREVENT ANYONE STUMBLING OVER THEM OR
ANY DAMAGE CAUSED BY THE WHEELS OF INSTRUMENT TROLLEYS. THE SAME CAUTION APPLIES
TO THE STORAGE OF THE LEADS.
The patient cable should not be exposed to excessive mechanical stress. Whenever
disconnecting the leads, hold the plugs and not the cables.
The cable can be wiped with soapy water. Sterilization, if required, should be done with
gas only and not with steam. To disinfect, wipe the cable with hospital standard
disinfectant.
Cleaning the Thermal Printhead of the internal printer
If the printer is used a lot, a residue of printers ink (from the grid on the paper) can build
up on the printhead. This can cause the print quality to deteriorate. We recommend
therefore that every month the printhead is cleaned with alcohol as follows:
Remove the paper tray. The thermal printhead is found under the paper tray release
catch. With a tissue dampened in alcohol, gently rub the printhead to remove the ink
residue. If the printhead is badly soiled, the colour of the paper grid ink (i.e. red or
green) will show on the tissue.