Reference Manual
300550EN, Rev AA
December 2010
6-3
Rosemount 2240S
Section 6. Service and Troubleshooting
4. Choose Predefined if you like to view a basic selection of registers.
Choose the All option if you want to select a certain range of registers by
your own choice. You can specify a start value in the Start Register input
field, and the total number of registers to be displayed in the Number of
Registers field (1-500). Up to 50 registers is recommended for a quick
update of the list
(1)
.
5. Click the
Read
button to update the
View Input/Holding Registers
window with fresh device data.
6.2.2
Editing Holding
Registers
Most Holding Registers can be edited simply by typing a new value in the
appropriate Value input field. Some Holding Registers (marked grey in the
Value column) can be edited in a separate window. In this case you can
choose from a list of options or you can change separate data bits.
For more information see the
Rosemount Raptor System Configuration
Manual (document no. 300510EN)
.
Predefined
All
(1) Note that Input Register data from the 2240S transmitter is temporarily stored in the Input
Register database of the 2410 Tank Hub. The Input Registers presented in TankMaster
WinSetup refer to the internal register area of the 2410. Therefore, for tank 1 you will have to
add 10000 to the 2240S internal register number as given by Table 6-4 in order to find the
register presented by WinSetup. For tank 2 (requires a 2410 with the multi-tank option) you
will have to add 12000, for tank 3 you will have to add 14000, and so on.
Содержание 2240S
Страница 2: ......
Страница 4: ......
Страница 58: ...Reference Manual 300550EN Rev AA December 2010 Rosemount 2240S 5 16 Section 5 Configuration Operation ...
Страница 84: ...Reference Manual 300550EN Rev AA December 2010 Rosemount 2240S A 6 Appendix A Reference Data ...
Страница 96: ...Reference Manual 300550EN Rev AA December 2010 Rosemount 2240S B 12 Appendix B Product Certifications ...
Страница 99: ......