Reference Manual
300550en, Rev AA
December 2010
Rosemount 2240S
www.rosemount-tg.com
Section 4
2240S Installation
Safety Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 4-1
Installation Considerations . . . . . . . . . . . . . . . . . . . page 4-2
Mechanical Installation . . . . . . . . . . . . . . . . . . . . . . page 4-3
Electrical Installation . . . . . . . . . . . . . . . . . . . . . . . . page 4-6
4.1 SAFETY
MESSAGES
Procedures and instructions in this section may require special precautions to
ensure the safety of the personnel performing the operations. Information that
raises potential safety issues is indicated by a warning symbol (
). Please
refer to the following safety messages before performing an operation
preceded by this symbol.
Failure to follow safe installation and servicing guidelines could result in death or
serious injury:
Make sure only qualified personnel perform the installation.
Use the equipment only as specified in this manual. Failure to do so may impair the
protection provided by the equipment.
Do not perform any service other than those contained in this manual unless you are
qualified.
WARNING – Substitution of components may impair Intrinsic Safety.
WARNING – To prevent ignition of flammable or combustible atmospheres, disconnect
power before servicing.
Explosions could result in death or serious injury:
Verify that the operating environment of the transmitter is consistent with the appropriate
hazardous locations certifications.
Before connecting a handheld communicator in an explosive atmosphere, make sure
the instruments in the loop are installed in accordance with intrinsically safe or
non-incendive field wiring practices.
Do not remove the gauge cover in explosive atmospheres when the circuit is alive.
High voltage that may be present on leads could cause electrical shock:
Avoid contact with leads and terminals.
Make sure the main power to the Rosemount 2240S is off and the lines to any other
external power sources are disconnected or not powered while wiring the gauge.
Содержание 2240S
Страница 2: ......
Страница 4: ......
Страница 58: ...Reference Manual 300550EN Rev AA December 2010 Rosemount 2240S 5 16 Section 5 Configuration Operation ...
Страница 84: ...Reference Manual 300550EN Rev AA December 2010 Rosemount 2240S A 6 Appendix A Reference Data ...
Страница 96: ...Reference Manual 300550EN Rev AA December 2010 Rosemount 2240S B 12 Appendix B Product Certifications ...
Страница 99: ......